Page 63 - DP Vol 20 No 3
P. 63

• Ultimately, screw was removed but it got crushed
                                                                               may be due to our previous attempts of removal.
                                                                               •  After  the  retrieval  of  the  broken  part  of  the
                                                                               abutment screw, a radiograph was taken to reconfirm
                                                                               its complete removal.
                                                                               • Prosthesis was placed again and patient is satisfied
                                                                               since then.

                                                                               DISCUSSION
                                                                               The current situation highlights some of the
                                                                               complexities associated with the use of implant
                                                                               components. In this case, the abutment screw
                                                                               fracture represents one of the rare complications
                                                                               encountered. The fracture is believed to have occurred
                                                                               due to the unnoticed loosening of the screw, which
                                                                               was  exacerbated  by  non-axial  movements  during
           Fig 4: Broken burs       Fig 5: Osstem ESR Screw Retrieval Kit      prosthetic loading, ultimately leading to the fracture.
                                                                                  Other possible reasons are:
                                                                               • Screw loosening can be a defect in the screw itself,
                                                                               • Screw may have been over-torqued on insertion,
                                                                               • The abutment may not have been fully seated when
                                                                               the torque was applied to the screw, resulting in
                                                                               increased stress.
                                                                               •  Several  techniques  of  removal  of  a  broken
                                                                               abutment screw have been documented, such as
                                                                               engaging an endodontic explorer or a straight probe
           Fig 6a-b: Radiographs while attempting retrieval                    and ultrasonic scaler.
                                                                                  In the described clinical situation, the broken
                                                     All the conventional methods   abutment screw could not be dislodged using
                                                  mentioned in literature for   an  explorer  or  an  ultrasonic  scaler  tip.  Despite
                                                  retrieval of broken screw were   employing all conventional methods in this case, we
                                                  tried.                       were unable to achieve the desired outcome.
                                                  •  An  ultrasonic  scaler  tip  was
                                                  then  inserted  into  the  implant   CONCLUSION
                                                  and was made to engage the   Implant screw breakage can occur for various reasons,
                                                  notch created.               and as clinicians, we must apply our expertise along
                                                  • Care was taken not to touch the   with the available tools and techniques to address
                                                  internal threads of the implant   each case individually. Many implant manufacturers
                                                  with the bur.                provide specialized screw retrieval systems. It is
           Fig 7: Remnants of Broken screw
                                                  •  The  ultrasonic  scaler  was   always advisable to use well-designed screw retrieval
                                                  started at a very low speed, and   kits, such as the one provided by Osstem Implant Pvt
                                                  mechanically the tip of the scaler   Ltd in the present case. n
                                                  was moved counter clockwise.
                                                  • Few burs which were available
                                                  online, surprisingly, they all broke
                                                  while attempting a retrieval.
                                                  •  Ultimately,  the  manufacturer
                                                  of the implant company was
                                                  contacted to know if they had any
                                                  prior experience with any dentist
           Fig 8: Radiograph after complete removal of   or not and surprisingly, they said   ABOUT THE AUTHOR
           broken screw                           that this was reported to them for
           the first time by any user. They have never encountered this query.                  (Prof.) Dr. Amit Bhardwaj
           • After not getting the support from the concerned implant company, I contacted      MDS, Ph.D, FICD,FPFA
           the Osstem implant company to seek their help and they contacted their Research &    Professor and Head,
           Development team for providing the solution.                                         Department of
                                                                                                Periodontology, Faculty
           • After a few days they provided Osstem ESR Kit for screw removal which had various   of Dental Sciences, SGT
           drills to provide easy solution for broken screw removal. The kit was easy to use and   University, Gurugram
           user friendly.